Description

The Olympus OmniScan SX Phased Array Ultrasonic Flaw Detector features a new streamlined software interface displayed on an 8.4 in. (21.3 cm) touch screen. A single-group and non-modular instrument, the OmniScan SX is easy to operate and cost-effective for less demanding applications.

The Olympus OmniScan SX touch screen offers a full-screen mode option that maximises visibility, essentially converting many menu functions into easy touch-screen operations. The intuitive interface provides smooth menu selection, zooming, gate adjustments, cursor movements, and text and value input. These, along with other premium integrated features, including easy-to-follow setup and calibration wizards, a rapid refresh rate for both the S-scan and A-scan displays, and a fast pulse repetition frequency (PRF), make the OmniScan SX a highly efficient inspection tool.

 

The OmniScan SX is fully compatible with the extensive portfolio of Olympus scanners, probes and accessories, as well as with its dedicated software companions, NDT SetupBuilder and OmniPC. Combined, the complete family of software and hardware contributes to a streamlined and efficient inspection workflow, from design and setup to acquisition and analysis.

The Omniscan SX unit comes supplied with 2 fully charged batteries and can be rented either on its own (to be used with your own probes and accessories) or with a range of probe and scanner kits which may be rented in a modular format. These include TOFD, Hydroform and Rexoform kits, together with chain scanners and glide scanners. Each of these kits has a range of accessories to suit most applications. Other probes and scanners may be available on request please contact us for details.

Olympus OmniScan SX Feature

  • Aperture: 16
  • Number of elements: 64
  • A streamlined software interface displayed on an 8.4 in. (21.3 cm) touch screen
  • Single-group and non-modular
  • Full-screen mode maximises visibility
  • 33% lighter and 50% smaller than MX2 model
  • Easy-to-follow setup and calibration wizards
  • Rapid refresh rate for the S-scan and A-scan displays
  • Fully compatible with the extensive portfolio of Olympus scanners, probes, accessories and software

Olympus OmniScan SX Applications

  • Phased array weld inspection
  • ToFD weld inspection
  • P/E weld inspection
  • General component inspection
  • Pressure vessel inspection
  • Composite inspection
  • Corrosion mapping inspection
  • Single group inspection
Olympus OmniScan SX Software Features
  • Interleave
  • Improved accessibility with more interactive menus
  • Curved geometry ray-tracing for longitudinal welds
  • Export C-Scan
  • Onboard compound scan
  • Analysis attenuation gain
  • New End-View with group merge
  • Optimized main menus and wizards
  • Scrolling layouts for easy interpretation
Olympus OmniScan SX Specification
 

Housing

dimensions (W x H x D)267 mm × 208 mm × 94 mm (10.5 in. × 8.2 in. × 3.7 in.)
Weight3.4 kg (7.5 lb) with battery
Data StorageStorage devicesSDHC card* or most standard USB storage devices.
*Lexar® brand memory cards are recommended for optimized results.
Data file size300 MB
I/O PortsUSB ports2 USB ports, compliant with USB 2.0 specifications
Audio alarmYes
Video outputVideo out (SVGA)
I/O LinesEncoder2-axis encoder line (quadrature, up, down, or clock/direction)
Digital input4 digital TTL inputs, 5 V
Digital output3 digital outputs TTL, 5 V, 15 mA maximum per output
Acquisition on/off switchYes, through configuration of a digital input
Power output line5 V, 500 mA power output line (short-circuit protected)
Pace input5 V TTL pace input
DisplayDisplay size21.3 cm (8.4 in.) (diagonal)
Resolution800 pixels x 600 pixels
Brightness600 cd/m2
Viewing anglesHorizontal: –80° to 80° Vertical: –60° to 80°
Number of colors16 million
TypeTFT LCD
Power SupplyBattery typeSmart Li-ion battery
Number of batteries1
Battery lifeMinimum 6 hours under normal operating conditions
Environmental SpecificationsOperating temperature range-10 °C to 45 °C
(14 ºF to 113 ºF)
Storage temperature range–20 °C to 60 °C (–4 ºF to 140 ºF) with battery
–20 °C to 70 °C (–4 ºF to 158 ºF) without battery
Relative humidityMax. 70% RH at 45°C noncondensing
Ingress protection ratingDesigned to meet requirements of IP66
Shockproof ratingDrop-tested according to MIL-STD-810G 516.6
Ultrasound Specifications (applies to OMNISX-1664PR)Connectors1 Phased Array connector: Olympus PA connector
2 UT connectors: LEMO 00
Number of focal laws256
Probe recognitionAutomatic probe recognition
Pulser/ReceiverAperture16 elements
Number of elements64 elements
PulserPA ChannelsUT Channel
Voltage40 V, 80 V, and 115 V95 V, 175 V, and 340 V
Pulse widthAdjustable from 30 ns to 500 ns; resolution of 2.5 nsAdjustable from 30 ns to 1,000 ns; resolution of 2.5 ns
Pulse shapeNegative square waveNegative square wave
Output impedance35 Ω (pulse-echo mode);
30 Ω (pitch- catch mode)
<30 Ω
ReceiverPA ChannelsUT Channel
Gain0 dB to 80 dB, maximum input signal 550 mVp-p (full-screen height)0 dB to 120 dB maximum input signal 34.5 Vp-p (full-screen height)
Input impedance60 Ω (pulse-echo mode);
150 Ω (pitch- catch mode)
60 Ω (pulse-echo mode);
50 Ω (pulse-receive mode)
System bandwidth0.5 MHz to 18 MHz (NOTE: The previously stated 0.6 MHz lower limit used a strict -3 dB attenuation for the cutoff frequency.)0.25 MHz to 28 MHz (–3 dB)
BeamformingScan typeSectorial or linear
Group quantity1
Data AcquisitionPA ChannelsUT Channel
Effective digitizing frequencyUp to 100 MHz100 MHz
Maximum pulsing rateUp to 6 kHz (C-scan)
Data ProcessingPA ChannelsUT Channel
Number of data pointsUp to 8,192
Real-time averagingPA: 2, 4, 8, 16UT: 2, 4, 8, 16, 32, 64
RectifierRF, full wave, half wave +, half wave –
Filtering3 low-pass, 3 band-pass, and 5 high-pass filters3 low-pass, 6 band-pass, and 3 high-pass filters (8 low-pass filters when configured in TOFD)
Video filteringSmoothing (adjusted to probe frequency range)
Data VisualizationA-scan refresh rateA-scan: 60 Hz; S-scan: 60 Hz
Data SynchronizationOn internal clock1 Hz to 6 kHz
On encoderOn 2 axes: from 1 to 65,536 steps
Programmable Time-Corrected Gain (TCG)Number of points16: One TCG (time-corrected gain) curve per focal law
Maximum slope40 dB/10 ns
AlarmsNumber of alarms3
Reference
Olympus OmniScan SX Brochure
Manufacturer Information

Customer Reviews